Operation
TCA-3 Temperature Monitoring System
English (EN)
31
1. Select Custom list from the Notify List field on general tab.
2. Add Email address and phone number to the list.
4.11.2 Trigger tab
The Trigger tab allows users to specify the conditions under which an alarm will be sounded.
The trigger mode of Auto-Rearm means that every time an alarm is triggered an alarm will be sounded. For
Example, if an alarm condition is triggered one after the other, two alarms will be heard.
The filter setting of more than 1 occurrence means that the alarm condition must exist for more than ONE
measurement. This prevents multiple alarms being sent for what really is a single alarm event.