Technical data
Eppendorf Reference
®
2
English (EN)
36
7.4 Test conditions
Test conditions and test evaluation match in compliance with ISO 8655, Part 6. Fine balance with
evaporation protection inspected by the office of weights and measures
Number of determinations:10
Use of water in accordance with ISO 3696
Inspection at 20 °C – 27 °C ±0.5 °C
Dispensing on the tube wall
7.5 Ambient conditions
Technical specifications subject to change.
Temperature range Relative humidity
Storage without transport
packaging
-5 °C — 45 °C 10 % — 95 %
Operating conditions 5 °C — 40 °C 10 % — 95 %